• 香蕉三级片,91香蕉视频污黄下载,香蕉污污视频,香蕉视频APP下载链接

    您好!歡迎訪問上海香蕉三级片科技有限公司網站!
    全國服務谘詢熱線:

    13390845525

    當前位置:首頁 > 資料下載 > 歐洲藥典-澄清度檢測規定

    歐洲藥典-澄清度檢測規定

    發布時間:2020/3/11      點擊次數:1353

    歐洲藥典-澄清度檢測規定

    歐洲藥典:

    Clarity and degree of opalescence of liquids

    VISUAL METHOD 一般方法

    Using identical test-tubes of colourless, transparent, neutral glass with a flat base and

    an internal diameter of 15-25 mm, compare the liquid to be examined with a reference

    suspension freshly prepared as described below, the depth of the layer being 40 mm.

    Compare the solutions in diffused daylight 5 min after preparation of the reference

    suspension, viewing vertically against a black background. The diffusion of light must

    be such that reference suspension I can readily be distinguished from water R, and that

    reference suspension II can readily be distinguished from reference suspension I.

    在內徑 15~25mm,平底,無色、透明、中性玻璃管中,加入等量的供試溶液與濁度標準液,

    使液位的深度都為 40mm,按如下所述方法進行比較。濁度標準液製備 5 分鍾後,以色散自然

    光照射濁度標準溶液和供試溶液,在黑色背景下從垂直方向觀察、比較澄清度或

    渾濁程度。色散自然光必須較容易區分濁度標準溶液Ⅰ與水,濁度標準溶液Ⅱ與濁度標準溶液Ⅰ

    A liquid is considered clear if its clarity is the same as that of water R or of the solvent

    used when examined under the conditions described above, or if its opalescence is not

    more pronounced than that of reference suspension I.

    如果供試溶液的澄清、透明程度與水相同,或者與所用溶劑相同,或者其澄清度不超過Ⅰ號濁度

    標準溶液,那麽可判定該溶液為澄清。

    Hydrazine sulfate solution. Dissolve 1.0 g of hydrazine sulfate R in water R and dilute to

    100.0 mL with the same solvent. Allow to stand for 4-6 h.硫酸肼溶液:取 1.0g 硫酸肼溶於水,加水稀釋至 100.0ml,靜置 4~6 小時。

    Hexamethylenetetramine solution. In a 100 mL ground-glass-stoppered flask, dissolve

    2.5 g of hexamethylenetetramine R in 25.0 mL of water R.

    烏洛托品(六亞甲基四胺)溶液 :在 100ml 容量瓶中,以 25.0ml 水溶解 2.5g 烏洛托品。

    Primary opalescent suspension (formazin suspension).

    To the examethylenetetramine solution in the flask add 25.0 mL of the hydrazine sulfate

    solution. Mix and allow to stand for 24 h. This suspension is stable for 2 months,

    provided it is stored in a glass container free from surface defects. The suspension must

    not adhere to the glass and must be well mixed before use.

    濁度標準貯備液:在存放烏洛托品溶液的 100ml 容量瓶中,加 25.0ml 的硫酸肼溶液。混合,

    靜置 24 小時,貯存在無表麵要求的玻璃容器中,可在 2 個月內使用。該濁度液不得黏附玻璃,

    用前必須充分搖勻。

    Standard of opalescence. Dilute 15.0 mL of the primary opalescent suspension to

    1000.0 mL with water R. This suspension is freshly prepared and may be stored for up to

    24 h.

    濁度標準原液:取濁度標準貯備液 15ml,加水稀釋、定容至 1000ml。該液臨用前製備,至多

    保存 24 小時。

    Reference suspensions. Prepare the reference suspensions according to Table 2.2.1.-1.

    Mix and shake before use.

    濁度標準液:由濁度標準原液與水按表 1-1 配製,即得。本液應臨用前配製。

    Table 1.-1Turbidity standard. The formazin suspension prepared by mixing equal volumes of the

    hydrazine sulfate solution and the hexamethylenetetramine solution is defined as a

    4000 NTU (nephelometric turbidity units) primary reference standard. Reference

    suspensions I, II, III and IV have values of 3 NTU, 6 NTU, 18 NTU and 30 NTU

    respectively. Stabilised formazin suspensions that can be used to prepare stable,

    diluted turbidity standards are available commercially and may be used after

    comparison with the standards prepared as described.

    濁度標準:乳光懸濁貯備液[硫酸肼溶液和烏洛托品(六亞甲基四胺)溶液以等量體積混合]定為

    4000NTU(比濁測定法的濁度單位)儲備液對照標準.濁度標準液 I, II, III 和 IV 相應的 NTU 值

    分別是 3 NTU, 6 NTU, 18 NTU 和 30 NTU。穩定的乳光懸濁貯備液可用於稀釋製備濁度標準,

    具有現實的商業價值,也可以與上述的標準化製備進行較。

    Formazin has several desirable characteristics that make it an excellent turbidity

    standard. It can be reproducibly prepared from assayed raw materials. The physical

    characteristics make it a desirable light-scatter calibration standard. The formazin

    polymer consists of chains of different lengths, which fold into random configurations.

    This results in a wide assay of particle shapes and sizes, which analytically fits the

    possibility of different particle sizes and shapes that are found in the real samples. Due

    to formazin’s reproducibility, scattering characteristics and traceability, instrument

    calibration algorithms and performance criteria are mostly based on this standard.

    因福馬爾肼具有一些香蕉三级片所希望的特性,所以它是一種非常優良的濁度標準物。它可以從被測原

    料中反複製備。具有所想要的光閃射校正標準化的物理特性。福馬爾肼聚合物由不同長度的鏈組

    成,他們可以折成各種形狀,應此可以分析不同大小和形狀的粒子。這一特性使得香蕉三级片可以對現實樣品中所具有的不同大小及性狀的粒子進行測定。由於。福馬爾肼具有可重複性、光散射性、

    可描繪性、儀器校準可算和操作標準化的特性,使其成為了濁度標準物。

    instrumental methods Introduction

    儀器方法簡介

    The degree of opalescence may also be determined by instrumental measurement of

    the light absorbed or scattered on account of submicroscopic optical density

    inhomogeneities of opalescent solutions and suspensions. 2 such techniques are

    nephelometry and turbidimetry. For turbidity measurement of coloured samples, ratio

    turbidimetry and nephelometry with ratio selection are used.

    該儀器是根據渾濁液和懸濁液亞顯微鏡光密度的不均一性來測量光的吸收或光的散射,即散射測

    濁法和透射測濁法。對於有色樣品的濁度測試法,要用到比率透射比濁法和可選擇比率的散射比

    濁法。

    The light scattering effect of suspended particles can be measured by observation of

    either the transmitted light (turbidimetry) or the scattered light (nephelometry). Ratio

    turbidimetry combines the principles of both nephelometry and turbidimetry.

    Turbidimetry and nephelometry are useful for the measurement of slightly opalescent

    suspensions. Reference suspensions produced under well-defined conditions must be

    used. For quantitative measurements, the construction of calibration curves is essential,

    since the relationship between the optical properties of the suspension and the

    concentration of the dispersed phase is at best semi-empirical.

    通過投射光(投射比濁法)或散射光(散射比濁法)來測量混懸粒子的光散射效能。濁度比率結

    合了透射比濁法和散射比濁法二者的原理。透射比濁法和散射比濁法用於測量具有輕微乳光的混懸液。必須使用在的條件下製得的標準混懸液。因為混懸液的光學性質與分散相的濃度之間的關

    係多是一個半經驗值,所以定量測定主要使用標準曲線法。

    The determination of opalescence of colored liquids is done with ratio turbidimeters or

    nephelometers with ratio selection, since color provides a negative interference,

    attenuating both incident and scattered light and lowering the turbidity value. The

    effect is so great for even moderately colored samples that conventional

    nephelometers cannot be used.

    因為溶液顏色會產生負幹擾,衰減入射光和散射光並降低濁度值,用比率透射濁度法和可選擇比

    率的散射濁度法測定有色溶液的乳光。對於正好適度的有色樣品,效果非常好,以至於常規的濁

    度儀不再使用。

    The instrumental assessment of clarity and opalescence provides a more discriminatory

    test that does not depend on the visual acuity of the analyst. Numerical results are

    more useful for quality monitoring and process control, especially in stability studies.

    For example, previous numerical data on stability can be projected to determine

    whether a given batch of dosage formulation or active pharmaceutical ingredient will

    exceed shelf-life limits prior to the expiry date.

    用儀器來判斷澄清度和乳光,試驗所提供的分辨能力更強,不再依靠分析者的視覺敏銳性來判斷。

    對於定性監控和過程控製,特別是穩定性研究,數字化結果更有用。例如,之前所得的關於穩定

    性的數字化資料用於判斷一個給定批號的劑量成分或活性藥物組分是否超過了貯存期限或者沒

    過有效期。

    NephelometryWhen a suspension is viewed at right angles to the direction of the incident light, the

    system appears opalescent due to the reflection of light from the particles of the

    suspension (Tyndall effect). A certain portion of the light beam entering a turbid liquid

    is transmitted, another portion is absorbed and the remaining portion is scattered by

    the suspended particles. If measurement is made at 90° to the light beam, the light

    scattered by the suspended particles can be used for the determination of their

    concentration, provided the number and size of particles influencing the scattering

    remain constant. The reference suspension must maintain a constant degree of

    turbidity and the sample and reference suspensions must be prepared under identical

    conditions. The Tyndall effect depends upon both the number of particles and their size.

    Nephelometric measurements are more reliable in low turbidity ranges, where there is a

    linear relationship between

    nephelometric turbidity unit (NTU) values and relative detector signals. As the degree

    of turbidity increases, not all the particles are exposed to the incident light and the

    scattered radiation of other particles is hindered on its way to the detector. The

    maximum nephelometric values at which reliable measurements can be made lie in the

    range of 1750-2000 NTU. Linearity must be demonstrated by constructing a calibration

    curve using at least 4 concentrations.

    散射比濁法當混懸液在垂直於入射光的方向觀察,因混懸液粒子產生的反射,係統出現乳光 (丁

    達爾效應).。進入一個渾濁液的光束,一部分被透過,一部分被吸收,剩餘部分被懸濁粒子散射。

    如果在與光束 90°的方向檢測,假如粒子數量和大小對散射的影響維持常數,可以用懸濁粒子的

    光散射來測定他們的濃度。照溶液的濁度必須保持不變,並且樣品和對照混懸液在一樣的條件下製備。丁達爾效應)依賴於粒子的大小和數量。在低濁度範圍,光散射濁度法更可靠,散射法濁

    度單位值和有關檢測器信號成線性。隨濁濁度的增加,不是所有的粒子都能暴露在入射光下的,

    並且在到達檢測器的途徑中,其他粒子的散射光被阻礙。一個可靠的測量所能測量的大散射濁度

    值是 1750-2000 NTU。必須用至少 4 個濃度構建標準曲線來證明線性。

    Turbidimetry

    The optical property expressed as turbidity is the interaction between light and

    suspended particles in liquid. This is an expression of the optical property that causes

    light to be scattered and absorbed rather than transmitted in a straight line through the

    sample. The quantity of solid material in suspension can be determined by the

    measurement of the transmitted light. A linear relationship between turbidity and

    concentration is obtained when the particle sizes are uniform and homogeneous in the

    suspension. This is true only in very dilute suspensions containing small particles.

    Linearity between turbidity and concentration must be established by constructing a

    calibration curve using at least 4 concentrations.

    透射濁度法在液體中懸濁粒子和光之間存在相關性,這一光學特性表示為濁度。表示的是光在直

    線方向上發生的散射和吸收,而不是光直線通過樣品的透射光學特性,通過測量透射光來測定混

    懸液中固體物質的量。當混懸液中粒子的大小均一且性質相同,可獲得濁度和濃度之間的線性關

    係。僅僅在很稀的含有少量粒子的混懸液中,才可實現線性。必須使用至少 4 個濃度構建標準

    曲線來證明濁度和濃度間呈線性。

    Ratio Turbidimetry

    In ratio turbidimetry the relationship of the transmission measurement to the 90°

    scattered light measurement is determined. This procedure compensates for the lightthat is diminished by the colour of the sample. The influence of the colour of the

    sample may also be eliminated by using an infrared light-emitting diode (IR LED) at 860

    nm as the light source of the instrument. The instrument’s photodiode detectors receive

    and measure scattered light at a 90° angle from the sample as well as measuring the

    forward scatter (light reflected) in front of the sample along with the measurement of

    light transmitted directly through the sample. The measuring results are given in

    NTU(ratio) and are obtained by calculating the ratio of the 90° angle scattered light

    measured to the sum of the components of forward scattered and transmitted light

    values. In ratio turbidimetry the influence of stray light becomes negligible.

    Nephelometers are used for measurements ofthedegreeofopalescenceof;比率透射比濁

    法測定的是透視光的測量和 90°方向上;

    Table 2.2.1.-2

    INSTRUMENTAL DETERMINATION OF OPALESCENCE

    乳光的儀器測定

    Requirements in monographs are expressed in terms of the visual examination method

    with the defined reference suspensions. Instrumental methods may also be used for

    determining compliance with monograph requirements once the suitability of the

    instrument as described below has been established and calibration with reference

    suspensions I-IV and with water R or the solvent used has been performed.

    乳光的儀器測定在用準確的參比混懸液定義可見方法時已表明了要求。一旦後麵所規定的建立了

    儀器的適應性,並用參比混懸液 I-IV 和水或使用的溶劑進行校正,文中的方法也使用於儀器校

    正。Apparatus. Ratio turbidimeters or nephelometers with selectable ratio application use

    as light source a tungsten lamp with spectral sensitivity at about 550 nm operating at a

    filament colour temperature of 2700 K, or IR LED having an emission maximum at 860

    nm with a 60 nm spectral bandwidth. Other suitable light sources may also be used.

    Silicon photodiodes and photomultipliers are commonly used as detectors and record

    changes in light scattered or transmitted by the sample. The light scattered at 90 ±

    2.5° is detected by

    the primary detector. Other detectors are those to detect back and forward scatter as

    well as transmitted light. The instruments used are calibrated against standards of

    known turbidity and are capable of automatic determination of turbidity. The test

    results expressed in NTU units are obtained directly from the instrument and compared

    to the specifications in the individual monographs.

    儀器:使用可選擇的比率濁度計和濁度計時,用鎢燈作光源,在 2700K 的譜線標記溫度時,鎢

    燈在大約 550nm 處有特殊選擇性,或者用在 860nm 處有大發射並且有 60nm 光譜寬度的紅外

    發光二級管。也可以使用其他合適的光源。常用矽製光電二極管和光電倍增管作檢測器,並記錄

    因樣品產生的光散射或光透射的改變。主要檢測器檢測在 90 ± 2.5°方向上的光散射。其他的檢

    測器檢測朝後和朝前的光散射,就像測光透射一樣。使用的儀器用已知濁度的標準溶液來校正,

    並能夠自動測定濁度。從儀器上直接獲得用 NTU 單位表示的測定結果,並且,在個別文中與規

    定進行比較。

    Instruments complying with the following specifications are suitable.

    根據後麵的說明使用儀器— Measuring units: NTU. NTU is based on the turbidity of a primary reference standard

    of formazin. FTU (Formazin Turbidity Units) or FNU (Formazin Nephelometry Units) are

    also used, and are equivalent to NTU in low regions (up to 40 NTU). These units are

    used in all 3 instrumental methods (nephelometry, turbidimetry and ratio turbidimetry).

    — 測量單位:NTU,NTU 根據的是福爾馬肼標準儲備液的濁度。液使用 FTU(福爾馬肼濁度單

    位)或 FNU(福爾馬肼散射測濁法單位)單位,在低濁度範圍內等於 NTU(大於 40NTU)。

    這些單位在散射測濁法、濁度法、比率濁度法,三種儀器方法中均可使用。

    — Measuring range: 0.01-1100 NTU.

    — 測量範圍:0.01-1100NTU

    — Resolution: 0.01 NTU within the range of 0-10 NTU, 0.1 NTU within the range of

    10-100 NTU, and 1 NTU for the range > 100 NTU. The instrument is calibrated and

    controlled with reference standards of formazin.

    — 分辨率:在 0-10NTU 範圍內分辨率為 0.01NTU,在 10-100NTU 範圍內分辨率為 0.1NTU,

    在>100NTU 範圍內分辨率為 1NTU。用福爾馬肼的參比標準校正和控製儀器。

    — Accuracy: 0-10 NTU: ± (2 per cent of reading + 0.01) NTU. 10-1000 NTU: ± 5

    percent.

    — 度:0-10 NTU±(2%測量讀數+0.01)NTU,10-1000 NTU: ± 5%

    — Repeatability: 0-10 NTU: ± 0.01 NTU. 10-1000 NTU: ± 2 per cent of the measured

    value.

    — 重複性:0-10 NTU± 0.01 NTU,10-1000 NTU± 2%的測量值

    — Calibration: with 4 reference suspensions of formazin in the range of interest.Reference suspensions described in this chapter or suitable reference standards

    calibrated against the primary reference suspensions may be used.

    — 校正:用在感興趣範圍內的 4 中福爾馬肼參比混懸液。可以用按本章規定的參比混懸液或合

    適的相對於參比混懸液儲備液標有刻度的參比標準來校正。

    — Stray light: this is a significant source of error in low level turbidimetric

    measurement;

    stray light reaches the detector of an optical system, but does not come from the

    sample; < 0.15 NTU for the range 0-10 NTU, < 0.5 NTU for the range 10-1000 NTU.

    — 雜散光:在低水平的濁度測定中,雜散光是主要的誤差來源。雜散光就是能到達光學係

    統的檢測器,但不是由於樣品而產生的光。0-10 NTU 的範圍內雜散光< 0.15 NTU,10-100NTU

    的範圍內雜散光< 0.5 NTU。

    Instruments complying with the above characteristics and verified using the reference

    suspensions described under Visual method may be used instead of visual examination

    for determination of compliance with monograph requirements.

    符合上麵的特性,並用在可見方法下規定的參比混懸液進行校正的儀器,在這範圍內,這些儀器

    可替代可視檢查,這些檢查和文中要求一致。

    Instruments with range or resolution, accuracy and repeatability capabilities other than

    those mentioned above may be used provided they are sufficiently validated and are

    capable for the intended use. The test methodology for the specific substance/product

    to be analysed must also be validated to demonstrate its analytical capability. The

    instrument and methodology should be consistent with the attributes of the product to

    be tested.所提供的儀器的使用範圍、分辨率、準確度、重複性、容量及其它上麵提到的參數,這些是十分

    有效的並能夠預期使用。對於分析特殊的物質/產品,也必須進行試驗方法學驗證來說明其分析

    能力。儀器和方法學應該和測試樣品的特性一致。

    上海香蕉三级片科技有限公司可以提供專業的澄清度檢測設備,歡迎大家谘詢。


    文件下載    
    上海香蕉三级片科技有限公司
    地址:上海市浦東新區航川路18號
    郵箱:info@yh-tek.com
    傳真:021-58220558
    關注香蕉三级片
    歡迎您關注香蕉三级片的微信公眾號了解更多信息:
    歡迎您關注香蕉三级片的微信公眾號
    了解更多信息
    網站地圖